Please note: The following represents a summary of the most important information needed to operate the Blanketrol II Unit. It is not meant to replace the Manufacturer’s Operation Manual and all staff operating the system need to familiarize themselves with the Manual. Refer to Section 3, CSZ Operation Manual for complete operating instructions.
A. Collect all equipment and supplies near the infant’s bedside
strainer. If needed, carefully add distilled/sterile water.
C. For each set of connecting hoses, attach to both a return and an outlet coupling on the unit side panel. Connect the other ends of the hoses to the like-colored couplings on the blankets. Hoses should not be twisted.
D. Be sure that the unit power switch is OFF prior to inserting the 3-prong plug into a properly grounded hospital grade receptacle.
E. Blankets should be flat and two clamps OPEN to allow blankets to fill.
F. Turn power switch ON. Water will be circulated to the blankets.
G. Check that there are no leaks in the blankets.
H. After the blankets have filled, once again check the level of distilled/sterile water in the reservoir. If needed, carefully add distilled/sterile water.
I. Set the Celsius/Fahrenheit switch so that Celsius is displayed.
J. Pre-cool the blanket by first operating in the Manual Control Mode.
A. Collect all equipment and supplies near the infant’s bedside
- Blanketrol II unit with a probe adapter cable and 2 sets of connecting hoses.
- 1 Infant size (25” x 33”) Maxi-Therm Lite blanket
- 1 Adult size (25” x 64”) Maxi-Therm Lite blanket
- YSI 400 series single patient use esophageal Steri-Probe
- Distilled/sterile water
- IV pole or portable rack (such as a garment rack) and soft clips (such as soft clip clothes hangers) with which to suspend the water-filled adult size blanket
strainer. If needed, carefully add distilled/sterile water.
C. For each set of connecting hoses, attach to both a return and an outlet coupling on the unit side panel. Connect the other ends of the hoses to the like-colored couplings on the blankets. Hoses should not be twisted.
D. Be sure that the unit power switch is OFF prior to inserting the 3-prong plug into a properly grounded hospital grade receptacle.
E. Blankets should be flat and two clamps OPEN to allow blankets to fill.
F. Turn power switch ON. Water will be circulated to the blankets.
G. Check that there are no leaks in the blankets.
H. After the blankets have filled, once again check the level of distilled/sterile water in the reservoir. If needed, carefully add distilled/sterile water.
I. Set the Celsius/Fahrenheit switch so that Celsius is displayed.
J. Pre-cool the blanket by first operating in the Manual Control Mode.
- Press the TEMP SET switch.
- Press the Down arrow to change the SETPOINT display to 5°C.
- Press the MANUAL CONTROL switch. (Allow about 15 minutes to pre-cool the blanket prior to use.)
- Soften the probe prior to insertion by placing in warm water for a few minutes.
- Position the temperature probe in the lower third of the esophagus. Nasal placement of probe is preferred. Measure distance from nares to the ear to the sternum (minus 2cm) and mark with an indelible pen before inserting.
- Secure the probe by taping it to the infant’s nose. Probe position may be confirmed with the next routine x-ray.